삼우과학홈페이지        
  거래업체
 

회사소개취급품목기술자료샘플이미지공지사항

 

pco. flim camera

Fluorescence (Luminescence) Lifetime Imaging application
simplified with the pco.flim



 

OVERVIEW

 

 
   

pco. flim Camera -

Luminescence lifetime imaging camera

The pco.flim camera system is the first luminescence decay time microscopy camera using a two tap CMOS image sensor. Synchronized modulation of pixels and stimulated light enables the analysis of decay times in the range of 100 ps to 100 μs. A broad variety of trigger options allows the camera's integration in versatile fields of applications. Additional technical information is provided in our data sheet.

Watch our pco.flim animation video.

Detailed information about fluorescence lifetime imaging using the pco.flim camera can be found in our new white paper explanation.

 

 
 

Key features

 

 
   

100ps - 100μs lifetimes measurable
5kHz - 40MHz modulation frequencies
500kHz - 40MHz external modulation signals
modulation signal shape sinusoidal / rectangular
1008 x 1008 pixel resolution
frequency domain FLIM
USB 3.0 interface
39% quantum efficiency
1000:1 dynamic range
45 e- rms readout noise
90 fps frame rate (2 tap readout)
10 ns to 10 s selectable exposure times,
vibration-free water cooling
special software for measurement and analysis

 

 
 

Technical DATA

 

 
    >
parameter Unit pco.flim
resolution (hor x ver) pixel 1008 x 1008
pixel size (hor x ver) um 5.6 x 5.6
modulation frequency (out) 5 kHZ - 40 MHz
modulation frequency (in) 500 kHz - 40 MHz
peak quantum efficiency % 39
dynamic range A/D bit 14